Processor crossover racks should be cleaned on a _____ basis.


A) daily
B) weekly
C) biweekly
D) monthly

Correct Answer

verifed

verified

The determination of how consistently radiologists' image interpretations match patients' ultimate diagnoses is called _____.


A) exposure linearity
B) outcome analysis
C) reproducibility
D) uniformity

Correct Answer

verifed

verified

Processors should be monitored every day for _____.


A) sensitometric measurements
B) water and chemical temperatures
C) replenishment tank levels
D) all of the above

Correct Answer

verifed

verified

Quality control testing procedures should be performed on _____ basis.


A) a daily and weekly
B) a monthly
C) an annual
D) all of the above

Correct Answer

verifed

verified

Following a good processor QC program reduces the need for _____ maintenance.


A) daily
B) unscheduled
C) scheduled
D) preventive

Correct Answer

verifed

verified

Focal spot size is measured with _____.


A) a pinhole camera
B) a star pattern
C) a slit camera
D) all of the above

Correct Answer

verifed

verified

A positive beam limitation (PBL) device is involved in?


A) Collimation
B) Exposure control
C) Dose reduction
D) Protocol monitoring

Correct Answer

verifed

verified

A

A quality control program is for _____.


A) film processors
B) CT scanners
C) x-ray tubes
D) all of the above

Correct Answer

verifed

verified

A quality assurance program monitors _____.


A) films
B) imaging equipment
C) people
D) processors

Correct Answer

verifed

verified

The x-ray field must center accurately to within ____% of SID.


A) 2
B) 3
C) 0.5
D) 1

Correct Answer

verifed

verified

What is the testing device used to measure cassette screen-film contact?


A) dosimeter
B) wire-mesh
C) sensitometer
D) star pattern

Correct Answer

verifed

verified

The collimator must be accurately aligned to within ±_____% of the SID.


A) 2
B) 5
C) 10
D) 12

Correct Answer

verifed

verified

Comparing the mR output at 50 mA @ 0.2 sec,100 mA @ 0.1 sec,and 200 mA @ 0.05 sec is part of a test for _____.


A) reciprocity
B) reproducibility
C) linearity
D) uniformity

Correct Answer

verifed

verified

The minimum HVL for an x-ray machine operated at 70 kVp is?


A) 1.0 mm Al
B) 5.0 mm Al
C) 2.0 mm Al
D) 20 mm Al

Correct Answer

verifed

verified

C

Acceptance testing should never be done by a _____.


A) medical physicist
B) manufacturer's representative
C) QC technologist
D) radiologic engineer

Correct Answer

verifed

verified

It is acceptable to perform a(n) _____ test instead of measuring the focal spot size.


A) line pair
B) exposure linearity
C) half-value layer
D) wire mesh pattern

Correct Answer

verifed

verified

Most quality control testing on radiographic equipment is done _____.


A) annually
B) semiannually
C) monthly
D) weekly

Correct Answer

verifed

verified

The first step of the QA recommended by the Joint Commission on Accreditation of Healthcare Organizations (JCAHO) is to _____.


A) collect and organize data
B) take action to improve care
C) assign responsibility
D) assess and document actions

Correct Answer

verifed

verified

Medical facilities have quality control programs in place in order to _____.


A) have JAHCO approval and accreditation
B) ensure quality patient care
C) receive payment from insurance carriers
D) do all of the above

Correct Answer

verifed

verified

A QC program should include _____,routine performance monitoring,and maintenance.


A) acceptance testing
B) outcome analysis
C) calibration
D) personnel monitoring

Correct Answer

verifed

verified

A