Acceptance testing should never be done by a ______________________.


A) medical physicist
B) manufacturer's representative
C) QC technologist
D) radiologic engineer

Correct Answer

verifed

verified

Comparing the mR output at 50 mA at 0.2 sec, 100 mA at 0.1 sec, and 200 mA at 0.05 sec is part of a test for ________________.


A) reciprocity
B) reproducibility
C) linearity
D) uniformity

Correct Answer

verifed

verified

Quality control testing procedures should be performed on ______________ basis.


A) a daily and weekly
B) a monthly
C) an annual
D) All of the above

Correct Answer

verifed

verified

The collimator must be accurately aligned to within ±_____% of the SID.


A) 2
B) 5
C) 10
D) 12

Correct Answer

verifed

verified

All general purpose radiographic units must have a minimum of _____ mm Al filtration.


A) 1.5
B) 2.0
C) 2.5
D) 3.0

Correct Answer

verifed

verified

When a light in a viewbox is out, _____________________.


A) that light bulb must be replaced
B) all bulbs in the viewbox must be replaced
C) all the bulbs must be matched to adjacent viewboxes
D) All of the above must be done

Correct Answer

verifed

verified

A quality assurance program monitors __________.


A) films
B) imaging equipment
C) people
D) processors

Correct Answer

verifed

verified

What is the testing device used to measure cassette screen-film contact?


A) Dosimeter
B) Wire-mesh
C) Sensitometer
D) Star pattern

Correct Answer

verifed

verified

B

The minimum HVL for an x-ray machine operated at 70 kVp is _________________.


A) 1.0 mm Al
B) 5.0 mm Al
C) 2.0 mm Al
D) 20 mm Al

Correct Answer

verifed

verified

Probably the most important patient protection characteristic of a radiographic imaging system is ______________.


A) filtration
B) focal spot size
C) accurate collimation
D) high contrast resolution

Correct Answer

verifed

verified

It is acceptable to perform a(n) _________ test instead of measuring the focal spot size.


A) line pair
B) exposure linearity
C) half-value layer
D) wire mesh pattern

Correct Answer

verifed

verified

The x-ray field must center accurately to within ____% of SID.


A) 2
B) 3
C) 0.5
D) 1

Correct Answer

verifed

verified

Filtration is measured with _____________ measurements.


A) caliper
B) photometer
C) slit camera
D) half-value layer

Correct Answer

verifed

verified

Following a good processor QC program reduces the need for __________ maintenance.


A) daily
B) unscheduled
C) scheduled
D) preventive

Correct Answer

verifed

verified

A QC program should include _________________, routine performance monitoring, and maintenance.


A) acceptance testing
B) outcome analysis
C) calibration
D) personnel monitoring

Correct Answer

verifed

verified

Medical facilities have quality control programs in place in order to _________________.


A) have TJC approval and accreditation
B) ensure quality patient care
C) receive payment from insurance carriers
D) Do all of the above

Correct Answer

verifed

verified

D

A quality control program is for ____________________.


A) film processors
B) CT scanners
C) x-ray tubes
D) All of the above

Correct Answer

verifed

verified

A positive beam limitation (PBL) device is involved in __________________.


A) collimation
B) exposure control
C) dose reduction
D) protocol monitoring

Correct Answer

verifed

verified

Focal spot size is measured with ________________.


A) a pinhole camera
B) a star pattern
C) a slit camera
D) All of the above

Correct Answer

verifed

verified

The first step of the QA recommended by The Joint Commission (TJC) is to __________________.


A) collect and organize data
B) take action to improve care
C) assign responsibility
D) assess and document actions

Correct Answer

verifed

verified